Evaluation of flow rates for six disposable infusion devices
D F Veal1, C E Altman, B T McKinnon
1Georgia Health Resources, Atlanta, USA.
Abstract:
The accuracy of flow was studied for six disposable infusion devices: Eclipse (Block Medical), Homepump (Block Medical), Intermate (Baxter Healthcare), MedFlo (Secure Medical), ReadyMED (McGaw), and SideKick (I-Flow). Each infusion device had a preset delivery rate of 100 mL/hr. Either 48 or 50 units of each device were filled with 100 mL of 0.9% sodium chloride injection. Flow was measured in a 60-mL syringe set 100 cm above the infusion device. The volume infused was recorded over 60 minutes at 5-minute intervals. Of the six devices tested, Homepump demonstrated the least variability in flow. The mean percentage of initial volume infused at 60 minutes was 93.7% for the Homepump, 92.4% for the MedFlo, 91.9% for the SideKick, 88.7% for the ReadyMED, 87.3% for the Eclipse, and 83.4% for the Intermate. The Intermate was the only device with a mean overall infusion rate outside the manufacturer's specifications of +/- 15%. The SideKick's mean infusion rate exceeded the +/- 15% range during the first half of the infusion. None of six disposable infusion devices delivered 100% of the volume to be infused within 60 minutes. Of the devices tested, the Homepump demonstrated the least variability in flow and the most complete volume delivery.
Related Concept Videos
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
In Vitro Drug Dissolution: Compendial Testing Models II
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations