Related Experiment Video
Updated: Aug 19, 2026

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Delayed success following radiofrequency catheter ablation
1Department of Medicine, New York Hospital-Cornell University Medical College, NY 10021.
Abstract:
Two patients who underwent attempted radiofrequency catheter ablation of an atrioventricular accessory pathway experienced delayed abolition of pathway conduction. During the procedures there was transient block in the accessory pathway following multiple ablation attempts at closely spaced sites. Both patients showed evidence of preexcitation at the conclusion of the ablation session but neither showed evidence of accessory pathway conduction during a second electrophysiological study. These observations demonstrate that in patients in whom transient block of the accessory pathway is produced, continued evolution of radiofrequency energy lesions may result in the eventual success of an initially unsuccessful ablation session.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
05:43Determination of Continuity Index Values in Atrial Fibrillation Ablation with Proactive Esophageal Cooling
Published on: April 19, 2024