Adapter failure as a cause of pacemaker malfunction

B M van Gelder1, F A Bracke, M I el Gamal

  • 1Department of Cardiology, Catharina Hospital, Eindhoven, The Netherlands.

Pacing and Clinical Electrophysiology : PACE
|October 1, 1993
PubMed
Summary

A protruding adapter set-screw caused short-circuiting during pulse generator replacement, leading to pacemaker noncapture. This resulted in dangerously low telementered impedance readings (< 100 ohms).

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Left bundle branch pacing as an alternative to biventricular pacing for cardiac resynchronisation therapy.

Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation·2022
Same author

Comparison between laser sheaths, femoral approach and rotating mechanical sheaths for lead extraction.

Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation·2021
Same author

Initial experience, feasibility and safety of permanent left bundle branch pacing: results from a prospective single-centre study.

Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation·2021
Same author

What is the cause of P-wave undersensing in this CRT-D device?

Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation·2018
Same author

What is the cause of P‑wave undersensing in this CRT-D device?

Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation·2018
Same author

The Hemodynamic Effects of Different Pacing Modalities After Cardiopulmonary Bypass in Patients With Reduced Left Ventricular Function.

Journal of cardiothoracic and vascular anesthesia·2017

Area of Science:

  • Cardiovascular medicine
  • Biomedical engineering
  • Implantable electronic devices

Background:

  • Pulse generator replacement is a common procedure for cardiac rhythm management.
  • Adapters are sometimes used to connect older leads to new pulse generators with different connector types (e.g., VS1 and IS1 3.2-mm connectors).
  • Ensuring proper lead-adapter-generator connection is critical for device function.

Observation:

  • Two patients experienced pacemaker noncapture following pulse generator replacement.
  • In both cases, an adapter was used to connect a preexisting lead to a new pulse generator.
  • The adapter's set-screw was observed to protrude through the insulation sleeve.

Findings:

  • The protruding set-screw caused a short circuit with the pulse generator's metal casing.

Related Experiment Videos

  • During noncapture events, the telementered lead impedance was measured below 100 ohms.
  • This indicates a significant electrical fault within the implanted system.
  • Implications:

    • Improperly secured adapters pose a risk of device malfunction and patient harm.
    • Regular impedance monitoring is crucial for detecting lead integrity issues.
    • Device manufacturers and clinicians should be aware of potential risks associated with adapter use in lead revisions.