Related Experiment Videos
Quantitative radionuclide angiocardiography: detection and quantitation of left to right shunts
Abstract:
In 105 patients defection and quantitation of left to right shunts was performed using quantitative radionuclide angiocardiography. The radionuclide angiocardiograms were acquired and analyzed by a gamma camera interfaced to a digital computer system. Pulmonary to systemic flow (Qp/As) ratios were calculated by analysis of pulmonary time-activity histograms using a gamma variate model. All patients were studied with cardiac catheterization, left ventricular angiocardiography and radionuclide angiocardiography. The radionuclide method allowed precise detection and quantitation of left to right shunts with a Qp/Qs ratio of 1.2 to 3.0. There was good agreement between the Ap/As ratio calculated by oximetry at cardiac catheterization and radionuclide angiocardiography (r = 0.94). The information gathered with this nontraumatic method appears sufficiently reliable to be used in the management of patients.