A comparison of methods for calibrating parallel-plate chambers/

C S Reft1, F T Kuchnir, L A DeWerd

  • 1University of Chicago, Illinois 60637.

Medical Physics
|December 1, 1994
PubMed
Summary

Determining the gas constant (Ngas) for parallel-plate ionization chambers is crucial for electron beam calibration. This study found three methods for calculating Ngas yield consistent results within 1% when using appropriate correction factors.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Erratum: "AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology".

Medical physics·2023
Same author

Electron contamination in <sup>6</sup><sup>0</sup> Co gamma-ray beams.

Medical physics·2018
Same author

SU-E-T-431: Investigation of BrachyVision™ Acuros™ Using Varian Surface Applicators.

Medical physics·2017
Same author

A round-robin gamma stereotactic radiosurgery dosimetry interinstitution comparison of calibration protocols.

Medical physics·2015
Same author

A prototype, glassless densitometer traceable to primary optical standards for quantitative radiochromic film dosimetry.

Medical physics·2015
Same author

Determination of the intrinsic energy dependence of LiF:Mg,Ti thermoluminescent dosimeters for 125I and 103Pd brachytherapy sources relative to 60Co.

Medical physics·2014

Area of Science:

  • Medical Physics
  • Radiation Dosimetry

Background:

  • Parallel-plate ionization chambers are essential for electron beam calibration.
  • Determining the gas calibration factor (Ngas) for these chambers is a significant concern in dosimetry protocols.
  • Current protocols like AAPM TG 21 recommend user-performed comparisons, while calibration labs use 60Co.

Purpose of the Study:

  • To verify the consistency of three different methods for determining Ngas in parallel-plate ionization chambers.
  • To compare user-performed methods with traditional calibration laboratory procedures.

Main Methods:

  • Measured Ngas for six common parallel-plate chambers using three distinct techniques.
  • Included methods based on AAPM TG 21, in-air exposure, and in-phantom exposure.
  • Utilized matched ionization chambers and phantom materials, irradiating with a 22 MeV electron beam and performing 60Co calibrations.

Related Experiment Videos

Main Results:

  • All three methods for determining Ngas produced results within 1% of each other for the tested chambers.
  • The consistency was achieved by applying appropriate correction factors specific to the chambers and methods.
  • This demonstrates the reliability of the evaluated procedures for Ngas determination.

Conclusions:

  • The study validates that multiple methods for calculating Ngas in parallel-plate ionization chambers can achieve high consistency.
  • The findings support the use of these methods, provided appropriate correction factors are applied.
  • This contributes to more reliable electron beam dosimetry and calibration practices.