[Inadvertant intracardiac entrapment of a Swan-Ganz catheter]

A Klockgether-Radke1, J Rathgeber, H Lange

  • 1Zentrum Anaesthesiologie, Rettungs- und Intensivmedizin, Georg-August-Universität Göttingen.

Der Anaesthesist
|February 1, 1995
PubMed

Insights

A Swan-Ganz catheter became entrapped within the heart during cardiac surgery due to a misplaced suture. Surgical release of the suture allowed for safe catheter removal, highlighting a rare complication of cardiac procedures.

Area of Science:

  • Cardiology
  • Cardiac Surgery
  • Medical Device Complications

Background:

  • A 58-year-old male patient presented with coronary artery disease and aortic valve insufficiency, necessitating cardiac surgery.
  • The patient underwent single coronary bypass grafting and ascending aorta replacement using a composite graft.

Observation:

  • A Swan-Ganz catheter was inserted via the right internal jugular vein for intraoperative monitoring.
  • Post-operatively, the catheter could not be removed, prompting further investigation.

Findings:

  • Fluoroscopy revealed intracardiac entrapment of the Swan-Ganz catheter.
  • Rethoracotomy identified the catheter ensnared by a purse-string suture.

Implications:

  • This case highlights a rare but serious complication of Swan-Ganz catheter use in cardiac surgery.
  • Proper suture technique and vigilance are crucial to prevent catheter entrapment.
  • Prompt diagnosis and surgical intervention are necessary for safe catheter retrieval.