Reconstructed helical CT scans: improvement in z-axis resolution compared with overlapped and nonoverlapped
C J Kasales1, K D Hopper, D N Ariola
1Department of Radiology, Pennsylvania State University, Hershey 17033, USA.
Objective:
This in vitro study was designed to assess the z-axis resolving capabilities of reconstructed helical CT scans obtained with various imaging parameters versus those of conventional CT scans and the effect of decreasing slice index on the z-axis resolution of helical CT.
Materials And Methods:
A z-axis line-pair phantom was imaged using conventional nonoverlapped CT scans, conventional CT scans that overlapped by 50%, and helical CT scans with pitches of 1.0 and 1.5. All helical images were reconstructed at comparable slice indexes (image indexes of 2.0, 1.0, and 0.5 mm for pitch = 1.0, and image indexes of 3.0, 1.5, and 0.75 mm for pitch = 1.5). Midline coronal and sagittal reconstructed images were obtained to allow standardized visualization of line pairs. The reconstructed images were reviewed separately by 10 radiologists.
Results:
The overall z-axis resolution of reconstructed helical CT scans equaled or exceeded that of nonoverlapped conventional CT scans in all cases and equaled that of 50% overlapped conventional CT scans in 75% of cases. The 1.0-pitch helical sequences showed improved z-axis resolution with decreasing slice index. No statistically significant improvement in z-axis resolution could be determined by the observers for 1.5-pitch sequences with decreasing slice index.
Conclusion:
The use of helical CT with a pitch of 1.0 or 1.5 and an increased slice index can improve the z-axis resolution of reconstructed images when compared with nonoverlapped conventional CT and frequently equals the resolution of 50% overlapped conventional CT. This improvement in z-axis resolution should improve the appearance of reconstructed images (as used in CT angiography and three-dimensional imaging) by reducing partial volume artifacts while affording faster scanning at a reduced skin-surface radiation dose.