Improved detection of silent cardiac ischemia with a 12-lead portable microprocessor-driven real-time
R C Thompson1, D C Mackey, G E Lane
1Division of Cardiovascular Diseases and Internal Medicine, Mayo Clinic Jacksonville, FL 32224, USA.
Objective:
To compare a microprocessor-driven real-time 12-lead electrocardiographic monitoring device with Holter monitoring for detection of ischemia.
Design:
Electrocardiographic monitoring was conducted in 110 patients at bed rest or undergoing surgical procedures.
Material And Methods:
In three groups of patients, simultaneous monitoring with a 12-lead real-time device and a 2-channel Holter system was performed to detect ischemic episodes. The differences in the number of ischemic events and the total time of ischemia revealed by the two devices were analyzed statistically.
Results:
In patients with coronary artery disease, more ischemic ST-segment shifts were detected by the 12-lead device than by Holter monitoring (44 versus 16 events; P < 0.05). Total time of ischemia was also greater with the 12-lead device (879 versus 273 minutes; P < 0.05). Ischemia was detected by both techniques in 6 patients, only by the 12-lead device in 12, and only by Holter monitoring in 1. Neither device detected ischemia in control subjects. The 12-lead device had an advantage in detecting inferior ischemia, and it identified an additional 13 patients with unstable angina who had changes in T-wave polarity but did not exhibit ST-segment shifts.
Conclusion:
The 12-lead real-time electrocardiographic monitoring device is superior to Holter monitoring in detecting and facilitating real-time identification of myocardial ischemia in patients at bed rest.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Holter Monitor: 24-Hour Monitoring