The use of a contrast-detail test object in the optimization of optical density in mammography
K J Robson1, C J Kotre, K Faulkner
1Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, UK.
Abstract:
In mammography, it is important that the maximum amount of diagnostic information is obtained from each radiograph. One of the factors influencing image quality is the optical density of the film. In this study, a contrast-detail test object was used to establish the optimum optical density value on the basis of signal-to-noise ratio (SNR) for two mammographic film-screen combinations under particular processing conditions. The optimum optical density, taken as being the point at which the threshold contrast is a minimum varied little with detail size. The optimum optical density for the Fuji film-screen combination with Photosol chemistry was found to be higher than that for the Kodak film-screen combination and Kodak processing conditions studied. Although the actual values of optical density are specific to the fixed processing conditions encountered, the technique has more general applicability.