[Do steroid-eluting pacemaker electrodes permit programming at lower pacemaker pulse amplitudes?]

A Schuchert1, K H Kuck, T Meinertz

  • 1Abteilung für Kardiologie, Universitäts-Krankenhaus Eppendorf.

Deutsche Medizinische Wochenschrift (1946)
|April 28, 1995
PubMed
Summary

Steroid-eluting ventricular electrodes allow pacemakers to be routinely set at 2.5 V and 0.5 ms in 96% of patients. However, individual parameter optimization remains crucial for all patients receiving these advanced pacing leads.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Optimising access: safety and efficacy of venous closure devices in cryoballoon pulmonary vein isolation.

Frontiers in cardiovascular medicine·2025
Same author

Predictors of Atrial Fibrillation Recurrence after Catheter Ablation: Data from the German Ablation Registry.

Scientific reports·2017
Same author

[Cardiac arrest in a fitness trainer with apical hypertrophic cardiomyopathy associated with cor triatriatum sinister].

Medizinische Klinik, Intensivmedizin und Notfallmedizin·2017
Same author

[Brain infarction as initial manifestation of eosinophilic myocarditis].

Der Nervenarzt·2017
Same author

Safety and efficacy of the percutaneous transaxillary access for transcatheter aortic valve implantation using various transcatheter heart valves in 100 consecutive patients.

International journal of cardiology·2017
Same author

The effects of local steroid-elution on high-performance pacing leads.

Herzschrittmachertherapie & Elektrophysiologie·2016

Area of Science:

  • Cardiology
  • Biomedical Engineering
  • Medical Devices

Background:

  • Steroid-eluting ventricular electrodes are designed to improve pacing thresholds.
  • Routine pacemaker settings aim to reduce energy consumption and device size.

Purpose of the Study:

  • To evaluate the feasibility of routinely setting pacemaker impulses at 2.5 V with a 0.5 ms duration using steroid-eluting ventricular electrodes.
  • To determine the safety and efficacy of lower stimulation parameters in a patient cohort.

Main Methods:

  • A study involving 107 patients implanted with steroid-eluting ventricular electrodes.
  • Measurement of impulse duration thresholds at various voltage amplitudes (0.8 V, 1.6 V, 2.5 V) 3-12 months post-implantation.
  • Analysis of pacemaker settings with a 100% margin of safety.

Related Experiment Videos

Main Results:

  • 96% of patients (103 out of 107) could be successfully paced at 2.5 V and 0.5 ms.
  • Impulse duration thresholds were significantly lower at 2.5 V (0.08 +/- 0.05 ms).
  • Four patients required alternative stimulation parameters; thresholds remained stable except in one case.

Conclusions:

  • Steroid-eluting electrodes enable reliable ventricular stimulation at 2.5 V and 0.5 ms in the majority of patients.
  • Individualized determination of optimal stimulation parameters is still necessary.
  • These findings support the use of lower pacing output with steroid-eluting electrodes, potentially enhancing device longevity.