Related Experiment Video
Updated: Aug 9, 2026

Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
The usefulness of calibration verification and linearity surveys in predicting acceptable performance in graded
G Lum1, D W Tholen, D A Floering
1Brockton/West Roxbury Veterans Affairs Medical Center, Boston, MA 02132, USA.
Objective:
To assess the association between performance on graded chemistry surveys and evaluation of linearity and calibration in Linearity surveys.
Design:
Data from Linearity Surveys (LN series) and from routine comprehensive College of American Pathologists chemistry surveys (all series) were used to evaluate the hypothesis that laboratories with nonlinear or univerified calibration would have a greater likelihood of unacceptable performance on comprehensive chemistry surveys.
Results:
This study found that acceptable calibration verification evaluation is significantly related to acceptable rates for most analytes, including albumin, calcium, chloride, glucose, iron, magnesium, sodium, total bilirubin, uric acid, high-density lipoprotein cholesterol, triglycerides, alkaline phosphatase, alanine and aspartate aminotransferase, digoxin, gentamicin, phenobarbital, procainamide, and thyroxine.
Conclusion:
There is a consistent and strong relationship between calibration verification problems in the Linearity Surveys and failure rates in the College of American Pathologists chemistry surveys. Laboratories with poor calibration evaluations on Linearity Surveys have higher unacceptable rates on proficiency tests. Individual laboratories who were rated linear and whose calibration was verified by Linearity Surveys have lower unacceptable rates.
More Related Videos
Related Concept Videos
Accuracy and Precision
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Calibration Curves: Correlation Coefficient
Data Validation
Key parameters for method validation include:
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...