Cardiac output measurement by the thermodilution method: an in vitro test of accuracy of three commercially available
A Rubini1, D Del Monte, V Catena
1Institute of Human Physiology, University of Padova, Italy.
Objective:
To describe the accuracy and the reproducibility of the thermodilution flow measurements obtained using 3 commercially available cardiac output computers commonly used in intensive care units.
Design:
An experimental "in vitro" study. Twelve different values of control flow (Qctr) were measured (Qmsr) using 3 different cardiac output computers (Abbott Critical Care System, Oximetrix 3 SvO2/CO Computer, Baxter Oximeter/Cardiac Output Computer SAT-1; American Edwards Laboratories, 9520 A Cardiac Output Computer). Standard equipment and techniques were employed, taking account of the specific weight and heat of warm water relative to blood. In addition, separate sets of measurements were performed in order to investigate the effect on Qmsr of some variables which may influence the "indicator" loss (time for injection, depth of immersion of the catheter, temperature of the injected fluid).
Setting:
Our laboratory, inside the intensive care unit.
Measurements And Results:
The analysis of the linear regression of Qmsr versus Qctr (r values between 0.992 and 0.984; residual standard deviation values comprised between 0.24 and 0.49 l/min; intercepts and slopes not significantly different from identity line), the values of the percentage errors (PE = [Qctr-Qmsr].100/Qctr; PE mean values 7.9, 5.0 and 13.1), and those of the coefficients of variability (CV = standard deviation mean value, %; CV mean values 5.4, 5.8 and 4.6), show a good level of accuracy and reproducibility of the measurements. Our data confirm previously reported results. Furthermore, the cumulative effect of variables capable of influencing the "indicator" loss, even if corrected according to the "calculation constant" the manufacturers provide, was found to result in statistically significant changes of Qmsr.
Conclusion:
The accuracy and reproducibility of the automatic cardiac computers tested is sufficient for practical clinical purpose. It may also depend on the modality of injection of the cooling bolus, which may significantly influence the effective "indicator" losses.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Pipe Flowrate Measurement
The orifice meter is a simple,...