Characteristics of the ventilator pressure- and flow-trigger variables
1Pulmonary and Critical Care Section, UCI-Long Beach Medical Program, Veterans Affairs Medical Center, CA 90822, USA.
Intensive Care Medicine
|February 1, 1995
Summary
Comparing two ventilators, the Puritan Bennett 7200ae showed higher work of breathing with pressure-triggered continuous positive airway pressure (CPAP) than flow-triggered CPAP. Adding pressure support to the 7200ae reduced this difference.
Area of Science:
- Mechanical Ventilation
- Respiratory Physiology
Background:
- Pressure- and flow-triggering are key features in modern mechanical ventilators.
- Continuous positive airway pressure (CPAP) modes are widely used for respiratory support.
- Understanding ventilator triggering mechanisms is crucial for optimizing patient-ventilator synchrony.
Purpose of the Study:
- To compare the triggering characteristics of pressure- and flow-triggered CPAP between the Puritan Bennett 7200ae and Siemens SV 300 ventilators.
- To evaluate the impact of these triggering methods on the lung pressure-time product (total PTP) using a mechanical lung model.
Main Methods:
- Utilized a mechanical lung model to assess ventilator performance.
- Measured the total pressure-time product (PTP) for pressure- and flow-triggered CPAP on both ventilator models.
- Investigated the effect of applying 5 cmH2O pressure support.
Main Results:
- The Puritan Bennett 7200ae exhibited a greater total PTP with pressure-triggered CPAP compared to flow-triggered CPAP due to insufficient flow delivery.
- Total PTP was comparable between pressure- and flow-triggered CPAP in the Siemens SV 300.
- The Siemens SV 300 demonstrated lower total PTP than the Puritan Bennett 7200ae for both triggering methods.
- Applying minimal pressure support to the Puritan Bennett 7200ae eliminated the difference in total PTP between triggering modes.
Conclusions:
- Ventilator triggering mechanisms and specific ventilator models significantly influence the work of breathing, as indicated by total PTP.
- The Siemens SV 300 appears more efficient in delivering CPAP compared to the Puritan Bennett 7200ae.
- Further research is needed to determine the clinical implications of these findings on patient inspiratory muscle work.