Related Experiment Videos
[Binocular photometer for measuring light perception in diseases of the visual pathway]
Abstract:
In patients with a reduction in visual function because of optic nerve neuritis, light sensitivity is tested to detect differences between the right and left eye in combination with the swinging flashlight test and visual evoked potential (VEP). A binocular Zeiss microscope was modified to use integrated photometer-controlled binocular stimulus fields and a central light system for fixation and variation of the contrast (background illumination). For exact binocular measurements of light sensitivity on both eyes, electronically controlled test fields (visual angle 5 degrees-20 degrees) guarantees only macular or overfoveolar stimulation. In addition, splitting the light for each stimulus field from only one halogen lamp guarantees wave-length-independent testing. Patients with various retinal diseases and neuritis nervi optici pathology were examined using pattern VEP, testing of the absolute lower threshold of light sensitivity, and binocular comparison over the range of luminous density. This binocular optical photometer system allowed minimal differences to be identified in the present visual function and after the eye had recovered from optic nerve neuritis.