[An unusual series of complications in therapy with implantable cardioverter-defibrillator]

W Peters1, H Hopp, W R Bauer

  • 1Medizinische Klinik, Universität Würzburg.

Zeitschrift Fur Kardiologie
|April 1, 1995
PubMed

Insights

Implantable cardioverter-defibrillator (ICD) oversensing due to insulation defects caused inappropriate shocks. Early detection and replacement of faulty components are crucial for effective ICD therapy.

Area of Science:

  • Cardiology
  • Biomedical Engineering

Background:

  • A 48-year-old male with dilative cardiomyopathy and ventricular fibrillation history received an implantable cardioverter-defibrillator (ICD) in 1990.
  • Early battery depletion necessitated a generator exchange, followed by recurrent inappropriate shocks.

Observation:

  • Despite comprehensive diagnostics, the cause of inappropriate shocks remained elusive for months.
  • Repeated beepograms eventually revealed oversensing, leading to generator and transvenous sensing electrode replacement.

Findings:

  • An insulation defect in the sensing electrode, near the generator, was identified as the cause of oversensing.
  • Following replacement, the patient experienced appropriate shocks for presyncope and is now stable.

Implications:

  • Implantable cardioverter-defibrillator (ICD) therapy requires specialized centers due to its complexity.
  • Thorough investigation and component integrity are vital for preventing device malfunction and ensuring patient safety.