Related Experiment Video
Updated: Aug 18, 2026

Levator Auris Longus Preparation for Examination of Mammalian Neuromuscular Transmission Under Voltage Clamp Conditions
Published on: May 5, 2018
State-of-the-art potential clamp device for myelinated nerve fibres using a new versatile input probe
K H Bohuslavizki1, A Kneip, E Koppenhöfer
1Institute of Physiology, University of Kiel, Germany.
Abstract:
A potential clamp device for myelinated axons is presented which, for the first time, systematically optimizes all the methodological parameters that limit the reliability of ionic current measurements. A crucial step toward this end consisted in the development of a new input probe with broad-band, extremely low-capacitance characteristics. In combination with a simple-to-use compensation criterion for optimum positive feedback, based on plausible assumptions, and with additional technological improvements, it enables the measurement of ionic currents with a reliability never previously achieved. The consequences with respect to evaluation of several earlier findings are discussed.