Related Experiment Videos
[A case-report of pacemaker malfunction caused by dry pocket after replacement of pulse generator]
H Yoshida1, O Ohba, Y Yamamoto
1Department of Cardiovascular Surgery, Hiroshima City Hospital, Japan.
Abstract:
Pacemaker malfunction was attributed to the increasing in impedance to current flow caused by a pocket of air separating the anodal contact plate of a unipolar generator from the overlying connective tissue (so-called dry pocket). Lack of capture was noted just after replacement of pulse generator. The malfunction was permanently corrected by bediside infusion of saline to pacemaker pocket.