The oesophageal detector device. An assessment of accuracy and ease of use by paramedics
Related Experiment Video
Updated: Aug 3, 2026

Simultaneous Laryngopharyngeal and Conventional Esophageal pH Monitoring
Published on: December 14, 2020
1University of Maryland Medical Center, Baltimore 21201.
Abstract:
Accuracy, ease and speed of recognition of tracheal tube position were assessed using the oesophageal detector device in a series of 40 tests on 29 patients. A single blind method was used, with each paramedic performing a single test on each patient. The tests were randomly split between two groups consisting of those tests performed on the tracheal or oesophageal tube respectively. Of the 40 tests performed, 21 were on tracheal tubes and 19 on oesophageal tubes. Correct placement was diagnosed with 100% accuracy by the paramedics, none of whom had previously used the oesophageal detector device. Each paramedic also graded speed of recognition of position and ease of use of the device. Recognition of position was graded as instant in 37 out of 40 tests. Use of the oesophageal detector device by previously inexperienced paramedics has thus been shown to be accurate, rapid and easy to learn.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management