Can a drawover vaporizer be a pushover?
1Army, Cambridge Military Hospital, Aldershot, Hampshire.
Anaesthesia
|October 1, 1994
Abstract:
Bench testing was carried out to establish whether the vapour output from an OMV50 vaporizer, as used in the Triservice apparatus, differs according to whether the carrier gas is either drawn or pushed through the vaporizer. Results show that the differences in output concentration between the two modes were clinically insignificant.