Automated analytical systems for drug development studies. I--A system for the determination of drug stability
1Department of Pharmaceutical Chemistry, University of Kansas, Lawrence 66045.
Abstract:
An automated system consisting of a pH-stat, microdialysis sampling and a liquid chromatograph was assembled to measure the rate of rapid chemical reactions. 2',3',5'-Triacetyl-6-azauridine was used as a model compound to validate the performance of the automated system. Buffer catalysis was minimized by using a non-catalytic concentration of borate buffer along with a pH-stat to maintain the pH during the kinetic run. The microdialysis sampling technique permitted sample quenching and buffering of the solutions to a pH compatible with the LC column materials. The combination of microdialysis sampling and rapid LC analysis allowed reactions with a half-life of approximately 1 min to be sampled every 30 s. The rates of hydrolysis of the drug, measured at different conditions of temperature (37-70 degrees C) and pH (9.0-10.5) using the automated system, compared well with the previously determined values.