[Multicenter study of commercial kits for the determination of human IgG subclasses, using the ELISA technique]
Abstract:
We evaluated two commercially available sandwich type Elisa procedures for the measurement of IgG subclasses in human serum. Assay kits from The Binding Site and the Central Laboratory of the Netherlands Red Cross Blood Transfusion Service were tested in six laboratories. The performance of spectrophotometers, pipettes and dilutors were assessed at each center. Within-run precision was estimated according to the Valtec method (Société Française de Biologie Clinique). The overall coefficient of variation ranged from 4 to 50% depending on subclass and kit. We also evaluated the IgG2 and IgG4 specificity using four sera containing a monoclonal IgG2 or IgG4 (kappa or lambda type). Using total IgG and immunoelectrophoresis as a comparative technique, IgG2 kappa and IgG4 kappa were both underestimated, IgG2 lambda was overestimated while IgG4 lambda compared favorably. Polyclonal IgG subclasses were frequently overestimated in these sera suggesting cross-reactions with either monoclonal IgG or other polyclonal IgG. Antigen excess was investigated and not encountered with either kit. Our results demonstrate that these procedures are insufficiently accurate or precise for routine clinical use.