Direct and telemetered lead impedance

U M Ben-Zur1, S B Platt, J N Gross

  • 1Department of Medicine, Montefiore Medical Center, Bronx, New York.

Pacing and Clinical Electrophysiology : PACE
|November 1, 1994
PubMed
Summary

Telemetered lead impedance measurements (LIM) show a good correlation with direct measurements but are not identical. While stable in the first month, telemetered LIM can increase over the first year, necessitating a baseline measurement for accurate follow-up.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Histopathological correlation of ablation lesions guided by noncontact mapping in a patient with peripartum cardiomyopathy and ventricular tachycardia.

Pacing and clinical electrophysiology : PACE·2002
Same author

Role of reactive oxygen species in IL-1 beta-stimulated sustained ERK activation and MMP-9 induction.

American journal of physiology. Heart and circulatory physiology·2001
Same author

Sinus node recovery after 25 years of atrial flutter.

Pacing and clinical electrophysiology : PACE·2001
Same author

Mechanism of inhibition of matrix metalloproteinase-9 induction by NO in vascular smooth muscle cells.

Journal of applied physiology (Bethesda, Md. : 1985)·2001
Same author

Case report: is this SVT or VT? An exception to the rule.

Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing·2001
Same author

Internal transcardiac pericardiocentesis for acute tamponade.

The American journal of cardiology·2000

Area of Science:

  • Cardiology
  • Biomedical Engineering
  • Medical Device Technology

Background:

  • Lead impedance measurements (LIM) are crucial for assessing pacemaker lead function.
  • Telemetered LIM offers a non-invasive method for monitoring lead integrity over time.
  • Understanding the correlation and stability of telemetered LIM compared to direct measurements is essential for clinical practice.

Purpose of the Study:

  • To correlate telemetered lead impedance measurements (LIM) with direct LIM.
  • To assess the stability of LIM over time using both direct and telemetered methods.
  • To establish the reliability of telemetered LIM for long-term pacemaker lead monitoring.

Main Methods:

  • Direct and telemetered LIM were measured in 91 patients across 101 leads during implantation and replacement.

Related Experiment Videos

  • Direct LIM measurements were compared between initial implant and pulse generator replacement in 41 patients.
  • Telemetered LIM stability was assessed at immediate postoperative, 1-month, and 1-year follow-ups in 50 patients.
  • Main Results:

    • A good correlation was observed between direct and telemetered LIM for both atrial (r=0.58 acute, r=0.87 chronic) and ventricular leads (r=0.69 acute, r=0.68 chronic).
    • Mean direct LIM increased significantly in atrial leads (124 Ω, P < 0.001) but not ventricular leads (10 Ω, P = NS) between initial implant and replacement.
    • Telemetered LIM showed stability in the first month but tended to rise throughout the first year, particularly in atrial leads.

    Conclusions:

    • Telemetered LIM provides a valuable but imperfect correlation with direct LIM measurements.
    • Direct LIM at initial implantation is comparable to measurements at pulse generator replacement.
    • Establishing an immediate postoperative baseline for telemetered LIM is critical for accurate long-term monitoring due to potential impedance increases with lead maturation.