Interdevice reliability and validity assessment of the Nicholas Hand-Held Dynamometer
E Trudelle-Jackson1, A W Jackson, C M Frankowski
1School of Physical Therapy, Texas Woman's University, Dallas 75231.
The Journal of Orthopaedic and Sports Physical Therapy
|December 1, 1994
Summary
Two identical Nicholas Hand-Held Dynamometers (HHDs) showed low reliability when measuring hamstring strength. Clinicians should not compare measurements between different HHD devices.
Area of Science:
- Biomechanics
- Rehabilitation Engineering
- Musculoskeletal Research
Background:
- The Nicholas Hand-Held Dynamometer (HHD) demonstrates high reliability for repeated measurements using the same device.
- Clinical settings often utilize multiple HHD units for patient assessments.
- Assessing consistency between different HHD units is crucial for accurate and comparable patient data.
Purpose of the Study:
- To evaluate the interdevice reliability of the Nicholas Hand-Held Dynamometer (HHD).
- To determine the validity of the Nicholas HHD by comparing its measurements to a criterion standard.
- To investigate if measurements from two identical HHDs are consistent for hamstring strength assessment.
Main Methods:
- Thirty healthy female subjects participated in the study.
- Hamstring strength was measured using two identical Nicholas Hand-Held Dynamometers (Device A and Device B).
- Measurements were compared against a Kin-Com isokinetic dynamometer as the criterion standard for validity.
Main Results:
- The intraclass correlation coefficient (ICC) between the two HHDs was low (ICC = .58), indicating poor interdevice reliability.
- Pearson correlation coefficients showed strong agreement between each HHD and the Kin-Com (r = .85 for Device A, r = .83 for Device B).
- Analysis of variance revealed no significant difference between Kin-Com and Device A, but a significant difference between Kin-Com and Device B (p < .001).
Conclusions:
- Measurements obtained from two identical Nicholas Hand-Held Dynamometers may not be interchangeable.
- Clinicians should exercise caution when comparing hamstring strength data collected from different HHD units.
- Further research may be needed to understand and mitigate interdevice variability in HHD measurements.