Related Experiment Videos
Quantitative studies of different injection systems in capillary electrophoresis
S V Ermakov1, Zhukov MYu, L Capelli
1Keldysh Institute of Applied Mathematics, Russian Academy of Sciences, Moscow.
Abstract:
For a rigorous assessment of the precise amount of sample loaded, for quantitation purposes, different sample injection systems were evaluated with two commercially available units, Waters Quanta 4000 and Beckman P/ACE 2100. In the first system, sample introduction by hydrostatic means (i.e., placing the sample vial at some height, usually 10.1 cm, above the other capillary end) was evaluated. It was found that in this system there is a constant positive bias, i.e. the amount of sample loaded lies on a curve parallel and above the theoretically predicted loading curve. However, the excess of mass loaded was constant along the injection times explored (covering from 5 to 35 s) and, for a 75 microns capillary, was found to be of the order of +6 nL (above the expected injected value). Thus it is easy to correct for this sample bias. In the electrokinetic mode, a very good correlation between expected and predicted sample loads was obtained for both units. In the pressure system (by positive pressure from a nitrogen tank, Beckman unit) a substantial discrepancy was found between experimental and predicted values (13.5% overload). Since the manufacturer claims a constant pressure of 0.5 psi, i.e. 3447.5 Pa, it would appear that the injection pressure is higher than the given value.(ABSTRACT TRUNCATED AT 250 WORDS)