Choice of moving target for a string phantom: II. On the performance testing of Doppler ultrasound systems
1Medical Physics Department, Royal Infirmary, Edinburgh, UK.
Abstract:
A string phantom was used to test a duplex system using several filaments: braided silk, nylon and O-ring rubber. The estimated pulsatility index was identical for all filaments tested. Range gate registration measurements were identical for all filaments tested. Maximum velocity was underestimated in the region of a strong angular backscatter peak; this was the case for several of the braided silk filaments. This effect was not observed for O-ring rubber. The paper concluded that O-ring rubber is a suitable filament for the testing of several key Doppler quantities. Recalibration of the string phantom velocity indicator is necessary due to the large diameter of the O-ring rubber.
Related Concept Videos
Doppler Effect - II
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: