Automatic mechanical device to standardize active compression-decompression CPR
V Wenzel1, R S Fuerst, A H Idris
1Department of Anesthesiology, University of Florida College of Medicine, Gainesville.
Study Objective:
To develop an automatic mechanical device capable of performing active compression-decompression (ACD) CPR in laboratory animals.
Design:
A swine model was used to study standard and ACD CPR. One-minute periods of standard mechanical chest compressions were alternated with mechanical ACD CPR.
Setting:
University hospital laboratory.
Interventions:
A commercially available device that provided standard chest compressions only was modified to deliver ACD CPR.
Results:
The absolute difference in intrapleural pressure and tidal volume almost doubled during ACD CPR compared with that with standard CPR.
Conclusion:
The presence of a greater negative change in intrapleural pressure confirmed that active decompression of the chest had occurred and that the device was capable of performing ACD CPR. The device provides consistent rate, depth, force, and duty cycle.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation I: Adult
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management