Comparison of simultaneous versus sequential defibrillation pulsing techniques using a nonthoracotomy system

H H Hsia1, R B Kleiman, B T Flores

  • 1Clinical Cardiac Electrophysiology Laboratory, Hospital of University of Pennsylvania, Philadelphia.

Pacing and Clinical Electrophysiology : PACE
|July 1, 1994
PubMed
Summary

Sequential shock delivery pathways significantly lower defibrillation threshold (DFT) compared to simultaneous pathways in patients with implantable cardioverter defibrillators. This finding is crucial for optimizing device therapy.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Outcomes of Catheter Ablation of Ventricular Tachycardia in the Setting of Structural Heart Disease.

Current cardiology reports·2016
Same author

Ablation of ventricular tachycardia.

Herzschrittmachertherapie & Elektrophysiologie·2007
Same author

Clinical cardiac electrophysiology fellowship teaching objectives for the new millennium.

Journal of cardiovascular electrophysiology·2002
Same author

Occurrence of implantable defibrillator events in patients with syncope and nonischemic dilated cardiomyopathy.

The American journal of cardiology·2001
Same author

Effects of linear, irrigated-tip radiofrequency ablation in porcine healed anterior infarction.

Journal of cardiovascular electrophysiology·2001
Same author

Changes in local wall thickness correlate with pathologic lesion size following radiofrequency catheter ablation: an intracardiac echocardiographic imaging study.

Echocardiography (Mount Kisco, N.Y.)·2001

Area of Science:

  • Cardiology
  • Biomedical Engineering

Background:

  • Implantable cardioverter defibrillators (ICDs) are vital for managing life-threatening arrhythmias.
  • Optimizing shock delivery pathways is essential for effective defibrillation and patient safety.

Purpose of the Study:

  • To compare the defibrillation threshold (DFT) using simultaneous (SIML) versus sequential (SEQ) shock delivery pathways in patients with three-lead nonthoracotomy ICD systems.
  • To determine the optimal pathway for ICD therapy in this patient population.

Main Methods:

  • Sixteen patients with three-lead nonthoracotomy ICD systems were studied.
  • DFT was determined for both SIML and SEQ pathways, with shocks delivered in 2-4 J increments/decrements.
  • The first pathway tested was either SIML or SEQ, with the second pathway tested starting 2-4 J above the initial DFT.

Related Experiment Videos

Main Results:

  • Sequential pathways resulted in a significantly lower DFT compared to simultaneous pathways (14 +/- 6 J vs 18 +/- 6 J; P < 0.01).
  • In 7 of 16 patients, SEQ pathways lowered DFT by > 5 J, whereas only one patient experienced this with SIML.
  • No significant difference was observed in the time delay to shock delivery between the two pathways.

Conclusions:

  • Sequential shock delivery pathways are more effective in lowering the defibrillation threshold in patients with multilead nonthoracotomy ICD systems.
  • These findings have significant implications for selecting optimal shock pathways in ICD therapy.
  • SEQ pathways may offer improved defibrillation efficacy and potentially allow for lower energy outputs.