Phantom evaluation of the effect of film processing on mammographic screen-film combinations
1University of Sydney, Faculty of Health Sciences, School of Medical Radiation Technology, Lidcombe, NSW, Australia.
Abstract:
Mammographic image quality should be optimal for diagnosis, and the film contrast can be manipulated by altering development parameters. In this study phantom test objects were radiographed and processed for a given range of developer temperatures and times for four film-screen systems. Radiologists scored the phantom test objects on the resultant films to evaluate the effect on diagnosis of varying image contrast. While for three film-screen systems processing led to appreciable contrast differences, for only one film system did maximum contrast correspond with optimal phantom test object scoring. The inability to show an effect on diagnosis in all cases is possibly due to the variation in radiologist responses found in this study and in normal clinical circumstances. Other technical factors such as changes in film fog, grain and mottle may contribute to the study findings.