Improved temperature control of the Technicon tissue processor paraffin bath

W J Eisler, P H Polk

    Stain Technology
    |November 1, 1976
    PubMed
    Summary

    Modifications to the Auto-Technicon Tissue Processor Paraffin Bath ensure reliable temperature control. These enhancements maintain paraffin temperature within 0.25°C for extended periods, improving tissue processing consistency.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Multisample parallel array fraction collector: application in DNA alkaline elution studies.

    Analytical biochemistry·1988
    Same author

    Analytical techniques for cell fractions. XXIII. A stable thermal gradient device for heat denaturation studies on proteins.

    Analytical biochemistry·1978
    Same author

    Dual channel temperature recorder.

    Applied microbiology·1975
    Same author

    High-intensity vapor lamp for biological research.

    Applied microbiology·1972
    Same author

    Instrument design considerations and clinical applications of stable isotope analysis.

    Clinical chemistry·1971
    Same author

    Automated continuous culture of mammalian cells in suspension.

    Science (New York, N.Y.)·1970

    Area of Science:

    • Histotechnology
    • Laboratory Equipment Engineering

    Background:

    • The Auto-Technicon Tissue Processor Paraffin Bath is crucial for tissue embedding.
    • Maintaining stable paraffin temperature is essential for consistent tissue processing and diagnostic accuracy.
    • Existing units may experience temperature fluctuations affecting reliability.

    Purpose of the Study:

    • To describe modifications for the Auto-Technicon Tissue Processor Paraffin Bath.
    • To enhance the long-term reliability and temperature stability of the paraffin bath.
    • To ensure dependable paraffin temperature maintenance for histological applications.

    Main Methods:

    • Implementing a combination of electrical and mechanical modifications to the Auto-Technicon Tissue Processor Paraffin Bath.
    • Conducting extensive testing to evaluate the performance and reliability of the modified units.
    • Monitoring temperature stability over extended continuous operation periods.

    Main Results:

    • Modified units demonstrated long-term operational reliability.
    • Dependable temperature maintenance of the paraffin was achieved.
    • Temperature stabilization within a narrow range of 0.25°C was verified over many months.
    • Continuous operation did not compromise temperature consistency.

    Conclusions:

    • The described modifications significantly improve the Auto-Technicon Tissue Processor Paraffin Bath's performance.
    • Enhanced reliability and precise temperature control are key benefits.
    • These improvements contribute to more consistent and dependable tissue processing in histology labs.