Related Experiment Videos
Simple high-performance liquid chromatographic analysis of free primary amino acid concentrations in rat plasma and
D J Begley1, A Reichel, A Ermisch
1Biomedical Sciences Division, King's College London, UK.
Abstract:
The quantitation of 16 acidic, basic, small and large neutral amino acids was performed using 10-microliters sample aliquots of cisternal cerebrospinal fluid (CSF) and blood plasma of rats. The analytical technique is based upon a two-buffer HPLC system with fluorimetric detection of pre-column derivatized primary amino acids with o-phthaldialdehyde (OPA). A modification of a well established method, the power of the present technique comes from an improved resolution and sensitivity by installing a column heater adjusted to 43 degrees C and strictly reducing any contamination by background amino acids. The analysis is simplified by separating the amino acid derivatives with a linear buffer gradient and less time consuming by the use of a short analytical column with a higher flow-rate. Analytical precision, linearity of response and reproducibility were highly acceptable at both CSF and plasma concentrations of amino acids without changing any of the separation or detection parameters.