Prolonged patency with a new-design Teflon biliary prosthesis

U Seitz1, H Vadeyar, N Soehendra

  • 1Department of Endoscopic Surgery, University Hospital Eppendorf, Hamburg, Germany.

Endoscopy
|June 1, 1994
PubMed
Summary

A new Tannenbaum (TB) stent design without sideholes significantly reduced biliary endoprosthesis clogging compared to traditional pigtail (PT) stents. This innovation offers improved patency for patients with malignant biliary obstruction.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Endoscopic haemoclip ligation of pedunculated polyp before polypectomy.

World journal of gastroenterology·2016
Same author

A Simple Evaluation Tool (ET-CET) Indicates Increase of Diagnostic Skills From Small Bowel Capsule Endoscopy Training Courses: A Prospective Observational European Multicenter Study.

Medicine·2015
Same author

[Isolation and characterization of rapidly labelled high molecular RNA from freely suspended callus cells of parsley (Petroselinum sativum)].

Planta·2014
Same author

[Selective inhibition of the synthesis of AMP-rich RNA by α-amanitin in higher plant cells].

Planta·2014
Same author

[Transport of newly synthesized rRNA from the nucleus to the cytoplasm in freely suspended cells of parsley (Petroselinum sativum)].

Planta·2014
Same author

[Relationship between anthocyanin synthesis and activity of phenylalanine ammonia-lyase (PAL) in callus cells of Daucus carota].

Planta·2014

Area of Science:

  • Gastroenterology
  • Medical Devices
  • Surgical Innovation

Background:

  • Biliary endoprosthesis clogging is a long-standing challenge for endoscopists.
  • Existing plastic stents with sideholes are associated with an increased risk of clogging.

Purpose of the Study:

  • To compare the patency of a new Teflon straight stent without sideholes (Tannenbaum stent) against traditional Teflon pigtail stents with sideholes in patients with malignant distal biliary obstruction.

Main Methods:

  • A prospective, nonrandomized study involving 74 patients receiving pigtail (PT) stents and 55 patients receiving Tannenbaum (TB) stents.
  • Both stent types were 10 Fr Teflon, used for malignant distal biliary obstruction.
  • Successful stent insertion and reduction in serum bilirubin were primary procedural outcomes.

Related Experiment Videos

Main Results:

  • The Tannenbaum (TB) stent demonstrated significantly longer patency compared to the pigtail (PT) stent (p < 0.02).
  • Median survival was 100 days for PT stents and 130 days for TB stents.
  • The TB stent's performance was comparable to metallic self-expandable biliary stents.

Conclusions:

  • The Tannenbaum stent, lacking sideholes, shows significant promise in reducing biliary stent clogging and improving patency.
  • Further randomized controlled trials are needed to compare the TB stent with current plastic and metal stents.