Dynamic properties of body plethysmographs and effects on physiological parameters
1Department of Clinical Physiology, University Hospital of Lund, Sweden.
Abstract:
A model incorporating compliance, resistance, inertia, and the thermal time constant of plethysmographs is used to describe the effect of its dynamic properties on measured respiratory parameters. Using numerical simulation we studied the effect of distortion of flow signals from 13 infants in whom flow and esophageal pressure had been recorded. The distortion in amplitude, shape, and timing of the recorded flow patterns was dependent on the dynamic properties of the plethysmograph. For constant-volume "pressure" plethysmographs, errors of derived parameters such as compliance and resistance are very important if the thermal time constant is short. These errors are not corrected by calibrating the plethysmograph at the breathing frequency. Time correction of the flow signals in volume-displacement plethysmographs gives accurate results when the plethysmograph resistance and compliance are low. Overall, a volume-displacement plethysmograph with moderately high resistance of the flowmeter, corrected for internal pressure and inertia, gives the best possible results.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pathophysiology of Cardiac Performance
Measurement of Blood Pressure
Pulse
The pulse serves as a clinical indicator...