X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Modifying fluoroscopic views reduces operator radiation exposure during coronary angioplasty
M R Pitney1, R M Allan, R W Giles
1Eastern Heart Clinic, Prince Henry Hospital, Little Bay, Sydney, Australia.
Objectives:
This three-part study examined the feasibility of reducing operator radiation exposure during coronary angioplasty.
Background:
As case loads and complexity increase, some cardiologists are receiving increasing radiation scatter doses. Techniques to reduce this are therefore becoming more important.
Methods:
First, the determinants of the operator dose were assessed by measuring the differences in scatter dose with different camera views. The relative contribution of fluoroscopy as opposed to cine was then quantified. Finally, operators were provided with these data, and subsequent changes in technique were evaluated.
Results:
Left anterior oblique views resulted in 2.6 to 6.1 times the operator dose of equivalently angled right anterior oblique views. Increasing steepness of the left anterior oblique view also resulted in a progressive increase in operator dose, with left anterior oblique 90 degrees causing eight times the dose of left anterior oblique 30 degrees and three times that of left anterior oblique 60 degrees. In the 45 coronary angioplasty cases prospectively analyzed, fluoroscopy was found to be a greater source of total radiation than cine by a 6.3:1 ratio (range 1.1 to 15.8). Once operators were made aware of the importance of left anterior oblique fluoroscopy, there was a marked reduction in its use. When this was not feasible, there was a reduction in the steepness of the angulation. Left anterior oblique fluoroscopy during angioplasty of the left anterior descending and circumflex coronary arteries was reduced from 40% of total screening time to approximately 5%, and left anterior oblique angulation for fluoroscopy during angioplasty of the right coronary artery decreased from 43.6 degrees (+/- 9.1 degrees) to 29.4 degrees (+/- 2.2 degrees). Success rates (90% vs. 89%) and screening times (19.5 vs. 20.7 min) remained unchanged in 200 coronary angioplasties performed after the study. Average operator radiation dose (measured by radiation badges worn under lead at waist level) was reduced from 32.6 to 14.3 microSv/operator per week despite a slight increase in case load.
Conclusions:
Fluoroscopy is the major source of total radiation exposure during coronary angioplasty, with left anterior oblique views providing the highest dose. Modification of views is feasible and will result in significant reduction of operator radiation dose.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging