Discrepancy in HbA1c measurements performed at different local laboratories and at a selected central reference

O N Jensen1, N de Fine Olivarius, P H Petersen

  • 1Department of Clinical Chemistry, Odense University Hospital, Denmark.

Insights

Local HbA1c measurements showed significant bias compared to central labs, impacting patient care. A bias of 0.25% HbA1c is recommended to ensure accurate hemoglobin A1c results.

Area of Science:

  • Clinical Chemistry
  • Laboratory Medicine
  • General Practice

Background:

  • Glycated hemoglobin (HbA1c) is crucial for diabetes management.
  • Variability in HbA1c measurements between local and central laboratories can affect clinical decisions.
  • A general practice intervention study highlighted discrepancies in patient HbA1c results.

Purpose of the Study:

  • To quantify the bias in HbA1c measurements between local and central laboratories.
  • To determine acceptable bias levels for HbA1c testing in clinical practice.
  • To describe a control system for HbA1c method standardization and specificity.

Main Methods:

  • 66 patients in a general practice study had HbA1c measured locally and at a central reference laboratory.
  • Bias was calculated using the mean difference in HbA1c values.
  • A control system for method standardization and specificity was developed.

Main Results:

  • 97% of local HbA1c results were lower than central laboratory results.
  • A mean bias of -1.47% HbA1c was observed between local and central laboratories.
  • A bias of 0.5% HbA1c is considered acceptable, with 0.25% HbA1c being the maximum allowed.

Conclusions:

  • Significant bias in local HbA1c measurements poses challenges for practitioners, patients, and laboratories.
  • Establishing strict bias limits (≤0.25% HbA1c) is essential for reliable diabetes monitoring.
  • Implementing a robust control system is vital for standardizing HbA1c methods and ensuring specificity.