Influence of spacer device on drug delivery to young children with asthma

L Agertoft1, S Pedersen

  • 1Department of Paediatrics, Kolding Hospital, Denmark.

Insights

Spacer device volume is less critical for budesonide delivery in infants (10-25 months) than previously thought. Smaller spacers (under 750 ml) effectively deliver medication when used with a facemask.

Area of Science:

  • Pediatric Pulmonology
  • Respiratory Drug Delivery
  • Pharmacokinetics

Background:

  • Inhaled corticosteroids like budesonide are crucial for managing pediatric asthma.
  • Spacer devices improve medication delivery to the lungs, but optimal device characteristics for infants are debated.
  • Budesonide delivery varies significantly between different inhaler devices.

Purpose of the Study:

  • To compare the delivered budesonide dose from three different spacer devices in infants aged 10-25 months.
  • To assess the impact of spacer volume on budesonide delivery in young children.
  • To evaluate the relationship between minute ventilation and medication delivery efficiency.

Main Methods:

  • A randomized crossover study involving 20 children (10-25 months) receiving budesonide (200 mcg) via three spacers (Nebuhaler, Aerochamber, Babyspacer) with facemasks.
  • Budesonide dose deposited on a filter was measured to determine delivery efficiency.
  • Children's minute ventilation was recorded during a 30-second inhalation period.

Main Results:

  • The Aerochamber (140 ml) delivered significantly less budesonide (39.4 mcg) compared to the Nebuhaler (53.5 mcg) and Babyspacer (55.5 mcg).
  • All spacers were effectively emptied within 30 seconds by the children's minute ventilation (mean 5.0 L/min).
  • Spacer volume did not appear to be the primary determinant of budesonide delivery in this age group.

Conclusions:

  • Spacer volume is less critical for budesonide delivery in infants aged 10-25 months, provided volumes under 750 ml are used.
  • Effective medication delivery is achievable with various spacer sizes when used with facemasks and appropriate breathing patterns.
  • Further research may explore other factors influencing inhaled corticosteroid efficacy in infants.