Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient?
Related Experiment Video
Updated: Oct 12, 2026

Lavage-induced Surfactant Depletion in Pigs As a Model of the Acute Respiratory Distress Syndrome (ARDS)
Published on: September 7, 2016
J Boldt1, T Menges, M Wollbrück
1Department of Anesthesiology and Intensive Care Medicine, Justus-Liebig-University Giessen, FRG.
Objective:
Evaluation of continuous cardiac output monitoring based on the thermodilution technique in the critically ill.
Design:
Prospective clinical investigation.
Setting:
A surgical intensive care unit of a university hospital.
Patients:
Thirty-five critically ill patients (trauma and/or sepsis patients), who needed pulmonary artery catheterization. The patients were prospectively studied according to the following groups: a) patients with a heart rate of > 120 beats/min; b) those patients with a cardiac output of > 10 L/min; c) patients with a cardiac output of < 4.5 L/min; d) patients with a rectal temperature of > 39.0 degrees C; and e) patients with a pulmonary artery catheter inserted for > 4 days.
Interventions:
Therapies were carried out according to modern intensive care medicine protocols by physicians who were not involved in the study.
Measurements:
Cardiac output was monitored continuously using a new, modified pulmonary artery catheter. This catheter has a heating filament by which energy is transmitted to the circulating blood (modified thermodilution technique). A bedside microprocessor calculated cardiac output using a new algorithm. Standard bolus thermodilution technique (10 mL of ice-cold saline solution) was used to compare the continuous cardiac output measurement with the intermittent bolus cardiac output measurement.
Main Results:
A total of 404 pairs of intermittent (bolus) cardiac output and continuous cardiac output measurements were obtained from the 35 patients. The bias (mean difference between bolus cardiac output measurement and continuous cardiac output measurement) of all measurements was 0.03 +/- 0.52 L/min and the 95% confidence limit (mean difference +/- 2 SD) was -1.01/1.06 L/min. Also, continuous cardiac output measurement agreed closely with bolus cardiac output measurement (bias was 0.16 +/- 0.57 L/min in the cardiac output of > 10 L/min group; bias was -0.17 +/- 0.50 L/min for the cardiac output of < 4.5 L/min group). Increased temperature and prolonged length of stay did not influence the agreement of continuous cardiac output measurement with bolus cardiac output measurement (bias was 0.09 +/- 0.51 L/min in the > 39 degrees C rectal temperature group).
Conclusions:
Continuous monitoring of cardiac output using a modified pulmonary artery catheter with a heated filament has proven to be accurate and precise in the critically ill patient when compared with the "standard" intermittent bolus thermodilution technique. The continuous monitoring technique enhances our armamentarium for more intensive monitoring of these patients under a variety of circumstances.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...