Objective:
The measure of the precision and accuracy of three optic SvO2 systems in vitro (SAT-2 Baxter, oximetrix 3 Abbott, HEMOPRO-2 Spectramed in a wide range of saturations (24-85%) and hematocrit (20-40%).
Design:
In vitro comparison of the results given by a Co-Oximeter and the SvO2 systems.
Measurements And Results:
144 blood samples were tonometered with a variety of gas mixtures. Measurements of SvO2 were compared to those obtained by means of a Co-Oximeter, the difference (dSO2) between paired values were calculated. Precision was evaluated from confidence intervals (2SD) and accuracy was tested using dSO2 figures. As there was no difference in either precision or accuracy among catheters for a given system, data obtained from a given system were pooled. Mean precision was 14.2%, 8.6%, 9% for the Spectramed, Baxter and Abbott systems, respectively.
Conclusion:
The rather low reliability of the optic method should be borne in mind in patients with low SvO2 and/or low hematocrit.
More Related Videos
04:36Measurement of Tissue Oxygenation Using Near-Infrared Spectroscopy in Patients Undergoing Hemodialysis
Published on: October 2, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...