MRI scanner variability studies using a semi-automated analysis system

R J Hyde1, J H Ellis, E A Gardner

  • 1Ann Arbor Department of Veterans Affairs Medical Center, Department of Radiology, Kresge III, MI 48109-0553.

Magnetic Resonance Imaging
|January 1, 1994
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The Opioid Crisis: Prevalence and Markets of Opioids.

Forensic science review·2022
Same author

Forensic science educational programs (VI) - Programs in the US.

Forensic science review·2020
Same author

Mass-like peripheral zone enhancement on CT is predictive of higher-grade (Gleason 4 + 3 and higher) prostate cancer.

Abdominal imaging·2014
Same author

Renal remodeling in the first 24 months after abdominal radiotherapy.

Practical radiation oncology·2014
Same author

Initial investigation of acoustic droplet vaporization for occlusion in canine kidney.

Ultrasound in medicine & biology·2010
Same author

Rapid 3-D imaging of contrast flow: application in a perfused kidney phantom.

Ultrasound in medicine & biology·2009
Same journal

Reconstruction of MRI from undersampled k-spaces of double-contrast volume acquisitions using deep neural networks.

Magnetic resonance imaging·2026
Same journal

Radiofrequency-induced heating safety of brain MRI scans at 7 T in the presence of a shoulder implant.

Magnetic resonance imaging·2026
Same journal

Incremental diagnostic value of microstructural time-dependent diffusion MRI in differentiating PCNSL from glioblastoma over conventional MRI.

Magnetic resonance imaging·2026
Same journal

Enhanced motion compensation for free-breathing dynamic contrast-enhanced MRI with GROG-facilitated bunch phase encoding and Golden angle radial sampling.

Magnetic resonance imaging·2026
Same journal

The allegory of the cave: 10 years of AI shadows in radiology.

Magnetic resonance imaging·2026
Same journal

Conversion of 3 T liver, spleen, pancreas, and kidney R2* measurements to 1.5 T R2* equivalents: Validation of a theoretical framework.

Magnetic resonance imaging·2026
See all related articles

The Parallel Rod Test Object (PRoTO) and its analysis program accurately assess MRI quality control, identifying key performance metrics like signal-to-noise ratio and distortion for monitoring system variability.

Area of Science:

  • Medical Imaging
  • Radiology
  • Magnetic Resonance Imaging (MRI)

Background:

  • MRI system quality control (QC) requires precise and accurate testing tools.
  • Existing QC methods may not fully capture long-term system performance variability.

Purpose of the Study:

  • To extensively test the precision, accuracy, and utility of the Parallel Rod Test Object (PRoTO) and its analysis program for MRI QC.
  • To identify reliable quantitative measures for assessing MRI system performance over time.

Main Methods:

  • Utilized the PRoTO and a semi-automated analysis program for quantitative MRI QC.
  • Acquired data from two MRI systems over 6-12 months, performing single scans for multiple QC measures.
  • Compared performance metrics derived from spin-echo and field-echo sequences.

Related Experiment Videos

Main Results:

  • The PRoTO and analysis program demonstrated reproducibility and accuracy in measuring MRI system parameters.
  • Signal-to-noise ratio (SNR) and distortion were found to be sensitive and valuable QC measures.
  • Resolution was less sensitive to parameter changes compared to SNR and distortion; spin-echo sequences yielded more stable results than field-echo.

Conclusions:

  • The PRoTO is a viable tool for comprehensive MRI QC, enabling timely assessment of long-term variability.
  • Monitoring SNR and distortion is recommended for MRI QC, while resolution may be less critical.
  • Long-term variability assessment is crucial for comparing MRI system performance and maintenance needs.