Mechanical thrombectomy with the Amplatz device: human experience

S M Tadavarthy1, P D Murray, S Inampudi

  • 1Minneapolis Heart Institute, Abbott Northwestern Hospital, MN 55407.

Journal of Vascular and Interventional Radiology : JVIR
|September 1, 1994
PubMed
Summary

The Amplatz thrombectomy device offers a rapid and effective method for mechanical clot removal in arteries and grafts. Early results show promise, particularly for synthetic bypass grafts, with potential for wider application.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

TIPS with a polytetrafluoroethylene-lined stent graft and associated haemolytic anaemia.

Gut·2008
Same author

Renal artery embolization with diluted hot contrast medium: an experimental study.

Journal of vascular and interventional radiology : JVIR·2001
Same author

Career advancement of men and women in academic radiology: is the playing field level?

Academic radiology·2000
Same author

The nitinol vascular occlusion plug: preliminary experimental evaluation in peripheral veins.

Journal of vascular and interventional radiology : JVIR·2000
Same author

Injury potential to venous valves from the Amplatz thrombectomy device.

Journal of vascular and interventional radiology : JVIR·2000
Same author

The road to success: factors affecting the speed of promotion of academic radiologists.

Academic radiology·1999

Area of Science:

  • Vascular Surgery
  • Interventional Radiology
  • Medical Devices

Background:

  • Thrombosis in arterial and synthetic bypass grafts poses significant clinical challenges.
  • Effective and rapid thrombus removal is crucial for restoring blood flow and limb salvage.

Purpose of the Study:

  • To evaluate the early clinical experience with the Amplatz thrombectomy device for mechanical arterial and graft thrombectomy.
  • To assess the safety and efficacy of this novel thrombectomy technique.

Main Methods:

  • Preliminary data from 14 patients treated with the Amplatz thrombectomy catheter were analyzed.
  • The device was used in arterial polytetrafluoroethylene grafts, native arteries, and for venous thrombosis.

Main Results:

Related Experiment Videos

  • Complete clot removal was achieved in 11 patients, with partial removal in three.
  • Mean thrombectomy time was 2 minutes 45 seconds, with minimal distal embolization.
  • Mechanical clot dissolution consistently caused hemolysis without adverse clinical effects.

Conclusions:

  • Mechanical thrombectomy using the Amplatz device is a new, effective, and rapid technique for thrombus removal.
  • The device shows particular promise for occluded synthetic femoral-to-popliteal bypass grafts.
  • Steerability and contralateral approach capability could enhance the device's broader applicability.