We report a case of a 15-yr-old North American Indian female with haemoglobin Hammersmith, scheduled for elective tonsillectomy, whose arterial oxygen saturation could not be reliably monitored perioperatively because of technical limitations of conventional dual wavelength pulse oximetry. The patient was chronically icteric. She had an atrial septal defect with a small L-->R shunt demonstrated by echocardiography. On arrival in the operating room pulse oximetry (Nellcor-Model N100) demonstrated a saturation of 45% whilst breathing room air. Her oxygen saturation increased to 60% whilst breathing 100% oxygen via a face mask. An arterial blood gas performed whilst breathing 100% oxygen revealed a PaO2 of 418 mmHg. Tonsillectomy was completed uneventfully under general anaesthesia. The pulse oximeter did not provide any clinically useful information throughout the case. In conclusion, conventional dual wavelength pulse oximeters cannot give an accurate estimate of oxygenation in patients with haemoglobin Hammersmith. Assessment of oxygenation in these patients requires alternative monitoring techniques.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:44Near-Infrared Spectroscopy During Reactive Hyperemia for the Assessment of Lower Limb Vascular Function
Published on: March 22, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Oxygen Transport in the Blood
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...