Accuracy of disposable blood pressure transducers used in the critical care setting
R H Bailey1, J H Bauer, J Yanos
1Department of Internal Medicine, University of Missouri-Columbia 65212.
Objectives:
To assess the initial accuracy, drift in calibration over several hours, and decline in accuracy over daily use of blood pressure transducers used in the critical care setting.
Design:
Prospective, criterion standard.
Setting:
Three critical care units in a university hospital.
Subjects:
Twenty-seven consecutively available high-pressure transducers were used to measure intra-arterial blood pressures, and 11 consecutively available low-pressure transducers were used to measure central venous and pulmonary arterial blood pressures.
Interventions:
High-pressure transducers were compared with a mercury column manometer, and low-pressure transducers were compared with a water column manometer at three pressure levels. Data were collected initially after installation of the pressure transducer, at 3 hrs, and at 6 hrs. Data collections were repeated on subsequent days.
Measurements And Main Results:
a) The majority of pressure transducers are accurate; b) there is no drift in the accuracy of pressure transducers over several hours; c) the accuracy of pressure transducers does not decline over subsequent days. Some transducers over-estimated standard pressures by 10 mm Hg and/or underestimated standard pressure by 17 mm Hg.
Conclusions:
The majority of pressure transducers used in the critical care setting are accurate, although there are some transducers that may compromise patient care. Blood pressure transducers should be checked against a standard manometer upon installation, and daily during use.
More Related Videos
04:54Invasive Hemodynamic Assessment for the Right Ventricular System and Hypoxia-Induced Pulmonary Arterial Hypertension in Mice
Published on: October 24, 2019
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure