An evaluation of the microbial retention performance of three ventilator-circuit filters
Objectives:
to determine and compare the microbial retention efficiency of 3 types of heat and moisture exchange (HME) filters.
Design:
randomised prospective study.
Setting:
Intensive Care Unit at the Middlesex Hospital and the microbiology laboratory of the Department of Academic Microbiology, University College London Medical School.
Measurements And Results:
An aerosol challenge to each filter using a suspension containing. S. marcescens and the bacteriophage MS2. In total 45 filters of 3 types were tested, 15 previously unused and 30 that had been used in ventilator circuits of 30 patients. The 3 types of filter generally had microbial retention efficiencies of > 99.99% for both S. marcescens and the bacteriophage MS2.
Conclusion:
The 3 types of HME filter were effective microbial filters and comparable in their microbial retention.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)