Related Experiment Videos
High-frequency rotational ablation following failed percutaneous transluminal coronary angioplasty
U Dietz1, R Erbel, H J Rupprecht
1Second Medical Clinic, Johannes Gutenberg University, Mainz, Germany.
Abstract:
Percutaneous transluminal coronary angioplasty (PTCA) failed in 29 of 1,150 patients (2.5%) after successful passage of the guide wire. The reasons for failure were inability to pass the lesion with a balloon in 28 patients and inability to dilate the lesion in 1 patient. In these patients (15 stenoses and 14 chronic occlusions) rotational ablation was performed. We were able to pass the burr through the lesion in all of them, resulting in a reduction of diameter stenosis from 87 +/- 15 to 51 +/- 18%. Rotational ablation alone was initially successful (stenoses reduction > 20% and residual stenoses < 50%) in 15 of 29 (52%) patients. Additional PTCA was performed in 21 of 29 (72%) patients, in 8 to optimize the initially successful result and in 13 because the outcome was unsatisfactory. After dilatation the diameter stenosis was reduced to 41 +/- 14% immediately after the procedure and to 36 +/- 13% at 24 hr control. Overall success was achieved in 21 of 29 (72%) patients immediately after the procedure and in 26 of 29 (90%) patients at 24 hr control. No acute major complications occurred. We conclude that rotational ablation can be used as a safe and effective alternative when PTCA is not successful.