Simple procedures can markedly enhance automated immunoassay performance

G Ward1, M White, P E Hickman

  • 1Division of Chemical Pathology, Princess Alexandra Hospital, Woolloongabba, Queensland, Australia.

American Journal of Clinical Pathology
|July 1, 1994
PubMed
Summary

Simple procedural changes to the Ciba Corning ACS:180 immunoassay system significantly improved analytic performance. This enhancement achieved near-optimal precision for key hormones, outperforming older isotopic immunoassay methods.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The invasion pore induced by <i>Toxoplasma gondii</i>.

bioRxiv : the preprint server for biology·2024
Same author

Views and experiences of healthcare practitioners supporting people with COPD who have used activity monitors: "More than just steps".

Respiratory medicine·2023
Same author

A Systematic Review of Inequalities in the Mental Health Experiences of Black African, Black Caribbean and Black-mixed UK Populations: Implications for Action.

Journal of racial and ethnic health disparities·2022
Same author

<i>What Are the Experiences of People with COPD Using Activity Monitors?: A Qualitative Scoping Review</i>.

COPD·2022
Same author

Corrigendum to "Bacterial mutagenicity screening in the pharmaceutical industry" [Mutat. Res. 752(2) (2013) 99-118].

Mutation research·2020
Same author

Seeded droplet microfluidic system for small molecule crystallization.

Lab on a chip·2020

Area of Science:

  • Clinical Chemistry
  • Immunoassay Technology

Background:

  • Optimal immunoassay performance theoretically requires identical interassay and within-run precision.
  • Existing automated immunoassay systems may have performance limitations.

Purpose of the Study:

  • To evaluate the impact of simple operational modifications on the analytic performance of the Ciba Corning ACS:180 automated immunoassay system.
  • To assess if procedural changes could achieve near-optimal precision for hormone assays.

Main Methods:

  • Implemented two straightforward changes to the operating procedures of the Ciba Corning ACS:180.
  • Assessed analytic performance using the interassay coefficient of variation (CV) according to National Committee for Clinical Laboratory Standards (NCCLS) protocols.
  • Evaluated six hormones: free tetraiodothyronine, thyroid-stimulating hormone (TSH), luteinizing hormone (LH), follicle-stimulating hormone (FSH), prolactin, and human chorionic gonadotropin (hCG).

Related Experiment Videos

Main Results:

  • Near-optimal analytic performance was achieved for four of six hormones (TSH, LH, prolactin, hCG).
  • Low hormone concentration interassay CVs were reported: free T4 (1.74 pM), TSH (0.033 mIU/L), LH (0.21 U/L), FSH (0.69 U/L), prolactin (5.03 mU/L), and hCG (1.52 mU/L).
  • The operational enhancements demonstrated improved analytic performance for all assessed hormones compared to previous isotopic immunoassays.

Conclusions:

  • Simple modifications to immunoassay operating procedures can substantially enhance analytic performance.
  • The Ciba Corning ACS:180 system, with procedural adjustments, offers improved precision for multiple hormone assays.
  • These enhancements provide a more accurate and reliable immunoassay system, surpassing older technologies.