How should diastolic blood pressure be defined during pregnancy?
S G Blank1, G Helseth, T G Pickering
1Cardiovascular Center, New York (NY) Hospital-Cornell University Medical Center 10021.
Hypertension (Dallas, Tex. : 1979)
|August 1, 1994
Summary
Disappearance of Korotkoff sounds is more accurate than muffling for measuring diastolic blood pressure in pregnancy. K2 analysis provides a reliable, noninvasive method for accurate blood pressure readings during gestation.
Area of Science:
- Obstetrics and Gynecology
- Cardiovascular Physiology
- Medical Instrumentation
Background:
- Accurate measurement of diastolic blood pressure in pregnancy is crucial but debated.
- Korotkoff sound phases IV (muffling) and V (disappearance) are commonly used auscultatory methods.
- Previous research established K2 analysis as a noninvasive, accurate method in nonpregnant subjects.
Purpose of the Study:
- To compare the accuracy of K2 analysis with auscultatory methods (muffling and disappearance) for diastolic blood pressure in pregnant women.
- To evaluate the reliability of muffling and disappearance phases in hypertensive and normotensive pregnancies.
Main Methods:
- 58 women (42 hypertensive, 16 normotensive) underwent 556 blood pressure evaluations.
- Simultaneous K2 analysis and auscultation by two observers (A1, A2) were performed.
- Agreement between observers and comparison of auscultatory phases with K2 diastolic pressure were analyzed.
Main Results:
- Muffling was inconsistently detected, with lower prevalence in hypertensive pregnancies (A1: 47.5%, A2: 37.3%) vs. normotensive (A1: 73.2%, A2: 66.7%).
- Observer agreement on muffling was low (112/348 instances).
- Disappearance of sound closely approximated K2 diastolic pressure in hypertensive subjects, with a slight underestimation in normotensive subjects.
Conclusions:
- Disappearance of Korotkoff sounds (phase V) is a more reliable indicator of diastolic blood pressure in pregnancy than muffling (phase IV).
- K2 analysis offers a precise, noninvasive alternative for diastolic blood pressure monitoring in pregnant individuals.
- Clinical classification impacts the detection of Korotkoff sound muffling, highlighting its limitations.