Performance characteristics of a whole-body PET scanner
T R DeGrado1, T G Turkington, J J Williams
1Duke University Medical Center, Durham, NC 27710.
Methods:
This study characterizes the performance of a newly developed whole-body PET scanner (Advance, General Electric Medical Systems, Milwaukee, WI). The scanner consists of 12,096 bismuth germinate crystals (4.0 mm transaxial by 8.1 mm axial by 30 mm radial) in 18 rings, giving 35 two-dimensional image planes through an axial field of view of 15.2 cm. The rings are separated by retractable tungsten septa. Intrinsic spatial resolution, scatter fraction, sensitivity, high count rate performance and image quality are evaluated.
Results:
Transaxial resolution (in FWHM) is 3.8 mm at the center and increases to 5.0 mm tangential and 7.3 mm radial at R = 20 cm. Average axial resolution decreases from 4.0 mm FWHM at the center to 6.6 mm at R = 20 cm. Scatter fraction is 9.4% and 10.2% for direct and cross slices, respectively. With septa out, the average scatter fraction is 34%. Total system sensitivity for true events (in kcps/(microCi/cc)) is 223 with septa in and 1200 with septa out. Dead-time losses of 50% correspond to a radioactivity concentration of 4.9 (0.81) microCi/cc and a true event count rate of 489 (480) kcps with septa in (out). Noise-equivalent count rate (NECR) for the system as a whole shows a maximum of 261 (159) kcps at a radioactivity concentration of 4.1 (0.65) microCi/cc with septa in (out). NECR is insensitive to changes in lower gamma-energy discrimination between 250-350 keV.
Conclusions:
The results show the performance of the newly designed PET scanner to be well suited for clinical and research applications.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET