Early postoperative increase in defibrillation threshold with nonthoracotomy system in humans
H H Hsia1, R L Mitra, B T Flores
1Clinical Cardiac Electrophysiology Laboratory, University of Pennsylvania Hospital 19104.
Pacing and Clinical Electrophysiology : PACE
|June 1, 1994
Summary
The defibrillation threshold (DFT) temporarily increases after implantable cardioverter defibrillator surgery but returns to normal levels within 4-6 weeks. This finding is crucial for optimizing post-implantation device programming.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Implantable cardioverter defibrillators (ICDs) are crucial for preventing sudden cardiac death.
- Assessing the stability of the defibrillation threshold (DFT) early after ICD implantation is vital for patient safety and device efficacy.
- Nonthoracotomy lead systems are increasingly used, necessitating an understanding of their DFT characteristics.
Purpose of the Study:
- To evaluate the stability of the defibrillation threshold (DFT) in patients early after the implantation of a multi-lead nonthoracotomy implantable cardioverter defibrillator system.
- To determine if the DFT changes in the early postoperative period and when it returns to baseline levels.
Main Methods:
- Fifteen patients with multi-lead nonthoracotomy ICD systems underwent DFT measurements at implant, 2-8 days postoperation, and 4-6 weeks later.
- DFT was defined as the lowest successful defibrillation energy, with shocks delivered in incremental steps.
- Analysis compared DFT values across the three time points.
Main Results:
- Defibrillation thresholds were significantly higher in the early postoperative period compared to at implant (22.7 +/- 7.0 J vs 16.9 +/- 3.9 J, P < 0.05).
- In patients assessed at all three periods, DFT increased postoperatively but returned to baseline levels by 4-6 weeks (16 +/- 7.1 J vs 16.4 +/- 3.9 J at implant, P = N.S.).
Conclusions:
- A transient rise in DFT is observed early after the implantation of multi-lead nonthoracotomy ICD systems.
- The DFT appears to stabilize and return to baseline within 4-6 weeks post-implantation.
- These findings have significant implications for the programming of energy output in ICD patients during the early post-implantation phase.