Patient and staff exposure to glutaraldehyde from KeyMed Auto-Disinfector endoscope washing machine
D A Lynch1, P Parnell, C Porter
1Centre for Digestive Diseases, General Infirmary at Leeds, Great Britain.
Abstract:
Activated glutaraldehyde (2%) is the recommended agent for disinfection of endoscopic equipment. Exposure to the disinfectant is associated with side-effects and guidelines have been set to avoid these complications. Endoscope washing machines are used to provide automated high level disinfection of endoscopes as well as reduce exposure to the irritant aldehyde. We report a design fault in an endoscope washing machine which results in patients and staff being exposed to activated glutaraldehyde. The Auto-Disinfector (KeyMed) comprises a washing chamber supplied by three separate reservoirs (detergent, disinfectant, and rinse water) via a common channel. After the first cycle endoscopes are processed using previously used detergent and rinse water. Rinse water glutaraldehyde concentrations were measured in four machines during routine endoscopy lists on at least two occasions and showed a progressive rise in glutaraldehyde concentration up to 0.1% after two to six cycles. This results in staff being exposed to glutaraldehyde present on the processed instrument and the disinfectant being pumped from the internal channels of the endoscope into direct contact with the gastrointestinal mucosa during endoscopy at concentrations of 200-1000 ppm. The present atmospheric limit for glutaraldehyde is 0.2 ppm. Skin and mucosal irritation occur at concentrations of 0.3 ppm and severe synovitis in experimental animals at 100 pm. Tongue swelling and bloody diarrhoea with characteristic mucosal histological changes have been reported in patients exposed to activated glutaraldehyde on inadequately rinsed equipment. To reduce this problem the rinse water should be changed after every cycle and the endoscope should be dried thoroughly before use.
More Related Videos
Related Concept Videos
Hand hygiene
Hand washing...
Handwashing I: Introduction and Types of Equipment
Hand wash basins in clinical areas should have faucets that can be turned on and off without using the hands; that is, they should be non-touch or lever-operated.
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Handwashing III: During the Procedure and Post-Procedure Steps