[Microbial contamination of wash water in automatic washing and disinfection machines]

A Kolch1, L Krizek, D Schoenen

  • 1Hygiene-Institut Universität Bonn.

Zentralblatt Fur Hygiene Und Umweltmedizin = International Journal of Hygiene and Environmental Medicine
|November 1, 1993
PubMed

Insights

Microbial contamination was found in all tested cleaning and disinfection machines at university clinics. Bacteriological objectionable tube materials in these medical device reprocessing machines promote bacterial growth, risking recontamination of treated tools.

Area of Science:

  • Microbiology
  • Materials Science

Context:

  • Medical device reprocessing is critical in healthcare settings.
  • Cleaning and disinfection machines are widely used in university clinics for medical tool treatment.

Purpose:

  • To investigate the cause of systemic microbial contamination in the water tube systems of cleaning and disinfection machines.
  • To identify factors contributing to bacterial growth within these machines.

Summary:

  • Bacteriological examination of water tubes from seven cleaning and disinfection machines revealed systemic contamination.
  • Bacterial growth was observed on the inner surfaces of tubes, attributed to the use of bacteriological objectionable materials, not solely ion-exchangers.
  • Sterile filtration did not prevent contamination, and recontamination of treated medical tools is a risk.

Impact:

  • Highlights the need for careful selection of materials in medical device reprocessing equipment.
  • Suggests that current cleaning and disinfection machine designs may pose a risk of microbial recontamination.
  • Underscores the importance of addressing material-related contamination pathways to ensure patient safety.