Related Experiment Videos
Ingrowth surfaces. Plasma spray coating to titanium alloy hip replacements
R B Bourne1, C H Rorabeck, B C Burkart
1University Hospital, University of Western Ontario, London, Canada.
Clinical Orthopaedics and Related Research
|January 1, 1994
Summary
Plasma spray coatings on titanium alloy total hip replacements (THRs) offer superior fatigue strength compared to other cementless devices. Clinical evidence shows these plasma spray THRs perform comparably to cemented implants.
Area of Science:
- Biomaterials Engineering
- Orthopedic Surgery
- Medical Device Technology
Background:
- Traditional cementless total hip replacements (THRs) utilize sintered bead or diffusion-bonded fiber metal ingrowth surfaces.
- These traditional methods can compromise the fatigue strength of titanium alloy implants, retaining less than 50% of the original strength.
- Plasma spray coatings present a promising alternative for enhancing implant performance and longevity.
Purpose of the Study:
- To evaluate the efficacy and safety of plasma spray coatings on titanium alloy THRs.
- To compare the performance of plasma spray cementless THRs against other cementless and cemented THR designs.
- To assess the fatigue strength retention and clinical outcomes associated with plasma spray coated THRs.
Main Methods:
- Analysis of laboratory data on fatigue strength retention in plasma spray coated titanium alloy implants.
- Review of clinical trial data comparing plasma spray cementless THRs (Mallory-Head) with sintered bead cementless THRs (PCA) and cemented THRs.
- Assessment of short to medium-term follow-up data for clinical outcomes and implant fixation.
Main Results:
- Plasma spray coatings allow titanium alloy implants to retain approximately 90% of their original fatigue strength.
- Laboratory and clinical evidence indicate that plasma spray cementless THRs perform as well as, or better than, other cementless implants.
- Clinical trials demonstrated the superiority of plasma spray cementless THRs over sintered bead designs and equivalent results to cemented THRs.
Conclusions:
- Plasma spray coatings are a safe and predictable option for cementless fixation of titanium alloy THRs.
- This coating technology offers significant advantages in preserving implant fatigue strength.
- Plasma spray THRs demonstrate favorable short to medium-term clinical outcomes, making them a viable alternative in hip replacement surgery.